HASO™3 32 Shack-Hartmann

HASO™3 32 Shack-Hartmann

HASO3 is the most recent evolution of our award winning HASO product line.  Based on our patented Shack-Hartmann technology, HASO3 wavefront sensors are easy to integrate and provide fast, accurate and reliable measurements by taking advantage of the HASO family's standard functionalities that customers have come to rely on, including absolute measurement, wide dynamic range, unequalled accuracy and insensitivity to vibration.

HASO3 32 HASO3 76-GE HASO3 128-GE
Aperture dimension 4.9 x 6.1 mm² 8.7 x 11.4 mm² 14.6 x 14.6 mm²
Number of microlenses 32 x 40 76 x 100 128 x 128
Refractive microlens technology standard square standard square
Tilt dynamic range >±3° (520λ) >±3°(1100λ) >±3°(1500λ)
Focus dynamic range - minimum
local radius of curvature
20 mm 15mm
Focus dynamic range-maximum NA >0.1
Repeatability (rms) < λ/200
Wavefront measurement accuracy
in relative mode (rms)1
λ/150
Wavefront measurement accuracy
in absolute mode (rms)2
λ/100
Tilt measurement sensitivity (rms) 3 µrad 1 µrad
Focus measurement sensitivity (rms) 10-3 m-1 5.10-4 m-1 2.5.10-4 m-1
Spatial resolution ~160 µm ~115 µm
Maximum acquisition frequency 50 Hz 11 Hz 7.5 Hz
Processing frequency (CPU 3Ghz, 512 Mb RAM) 20 Hz 10 Hz 5 Hz
Working wavelength range 350 - 1100 nm
Calibrated wavelength range 400- 600 nm, 500 - 700 nm, 630 - 900 nm, 800 - 1100 nm
Extended wavelength range see 76-GE/128-GE 400-700 nm, 500-900 nm, 650-1100 nm
Working temperature 15 - 30° C
Dimension / weight 75 x 62 x 68 mm / 510 g 115 x 51 x 60 mm / 400 g
Power supply 12 V / 6 W
Interface FireWire Giga Ethernet
1) Difference between the real wavefront and a reference wavefront obtained in similar conditions (10 λ of shift maximum).
2) Wavefront as seen by the wavefront sensor. Performance kept on the whole spectral range.

PDF Datasheet